| Temperature options | Standard room temperature Low temp 10 K to 350 K (optical CCR), 400 K (CCR)
 High temp room temperature to 1273 K (oven)
 | 
|---|
| Sample type | van der Pauw, Hall bar | 
|---|
| Maximum sample size | 10 mm × 10 mm × 3 mm | 
|---|
| Software | MeasureLINK™ software included to execute Hall measurements | 
|---|
| Gate bias option | Up to 100 V | 
|---|
| Legacy conversion | Full legacy 8400 replacement | Partial legacy 8400 replacement | 
|---|
| Resistance range options | Low-resistance Coming soon Standard resistance 0.5 mΩ to 10 MΩ
 High-resistance 10 MΩ to 200 GΩ
 | Standard resistance 10 mΩ to 10 MΩ High-resistance 10 MΩ to 200 GΩ
 | 
|---|
| Measurement type | DC field Hall measurement AC field Hall measurement
 | DC field Hall measurement FastHall measurement
 (AC electrical, DC field)
 | 
|---|
| Measurement-specific instrumentation | Base 776B and M81-SSM-6 + BCS-10 Standard resistance BCS‑10, VM‑10
 High-resistance VM‑10, VS‑10, CM‑10
 | Standard resistance M91 High-resistance M91-HR
 | 
|---|
| DC field options (nominal) | 4-inch magnet 1.69 T room temp and 0.91 T variable temp 7-inch magnet 2.28 T room temp and 1.54 T variable temp
 | 
|---|
| AC field range options (fixed-field, nominal) | Frequency: 50 mHz or 100 mHz 4-inch magnet 1.18 T (RMS) room temp and 0.63 T (RMS) variable temp
 7-inch magnet 1.20 T (RMS) room temp and 0.69 T (RMS) variable temp
 | None | 
|---|
| Resistivity | 10-5 to 105 Ω·cm | 10-3 to 105 Ω·cm | 
|---|
| Carrier concentration density | 8 × 102 to 8 × 1023 cm-3 | 8 × 102 to 8 × 1023 cm-3 | 
|---|
| Mobility range | 1 × 106 (DC field) and 10-3 to 106 (AC field) cm2/V s
 | 10-3 to 106 cm2/V s | 
|---|
| Multi-carrier analysis | QMSA-compatible (DC field only) | QMSA-compatible | 
|---|