Measurement
Number of units: 1
Input type: 2-lead, ground-referenced
Input resistance: 100 kΩ or 10 kΩ
Maximum operating input voltage: 60 V
Absolute maximum input voltage: 100 V — WARNING — voltages between 60 V and 100 V will not damage the instrument but could result in personal injury or damage to other instruments
Update rate: 5 rdg/s display; 30 rdg/s IEEE-488; 30 rdg/s serial
DC
DC resolution: To 5¾ digits
DC integrator capacitance: 1 µF nominal
DC input resistance: 100 kΩ 10 kΩ
DC ranges: 300 mVs, 30 mVs, 30 mVs, 3 mVs
DC resolution: 0.001 mVs, 0.0005 mVs, 0.0005 mVs, 0.0005 mVs
DC accuracy: Offset: ±10 µVs ±DC integrator drift; Gain: ±0.25% of reading (<10 Vs/s maximum rate of change)
DC minimum dΦ/dt: 20 µVs/min
DC maximum dΦ/dt: 60 Vs/s
DC integrator drift: ±1 µVs/min, 0.0004% FS/min on 300 mVs range (100 kΩ input resistance constant temperature environment)
DC peak
DC peak resolution: 4¾ digits
DC peak integrator capacitance: 1 µF nominal
DC peak input resistance: 100 kΩ 10 kΩ
DC peak ranges: 300 mVs, 30 mVs, 30 mVs, 3 mVs
DC peak resolution: 0.01 mVs, 0.001 mVs, 0.001 mVs, 0.001 mVs
DC peak minimum reading: 0.05 mVs, 0.005 mVs, 0.005 mVs, 0.005 mVs
DC peak accuracy: Offset: ±100 µVs ±DC integrator drift; Gain: ±5% of reading (<10 Vs/s maximum rate of change)
DC peak maximum dF/dt: 60 Vs/s
DC peak update rate: Reduces update rate to ¼ normal
AC
AC resolution: 4¾ digits (reduced to 3¾ digits on the 30 µVs range)
AC integrator capacitance: 0.1 µF nominal
AC input resistance: 100 kΩ
AC ranges: 30 mVs, 3 mVs, 300 µVs, 30 µVs
AC resolution: 0.001 mVs, 0.0001 mVs, 0.01 µVs, 0.01 µVs
AC minimum rdg: 3.000 mVs, 0.3000 mVs, 30.00 µVs, 3.00 µVs
AC frequency response: 2 Hz to 50 kHz
AC accuracy: ±1% rdg ±10 µVs (10 Hz to 10 kHz sinusoidal); ±5% rdg ±10 µVs (2 Hz to 50 kHz sinusoidal)
AC integrator drift: NA
AC peak
AC peak resolution: 3¾ digits
AC peak integrator capacitance: 0.1 µF nominal
AC input resistance: 100 kΩ
AC peak ranges: 30 mVs, 3 mVs, 300 µVs
AC peak resolution: 0.01 mVs, 0.001 mVs, 1 µVs
AC minimum reading: 0.01 mVs, 0.001 mVs, 5 µVs
AC peak accuracy: ±5% rdg ±10 µVs (10 Hz to 10 kHz sinusoidal); ±10% rdg ±10 µVs (2 Hz to 50 kHz sinusoidal)
AC peak update rate: Reduces update rate to ¼ normal
Front panel
Display type: 2-line × 20-character, vacuum fluorescent display
Display resolution: To ±5¾ digits
Display update rate: 5 rdg/s
Display units: Vs, MxN, WbN, VF, MxF, WbF, G, T, Wbcm, A, %
Units multipliers: p, n, µ, m, k, M, G
Annunciators: AC – AC input signal, DC – DC input signal, ∧ ∨ – Positive and negative peaks, R – Remote operation, :Alarm on
Keypad: 21 full travel keys
Interfaces
IEEE-488.2 capabilities: SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT0, C0, E1
Serial interface: RS-232C electrical, DA-9 connector, 9600 BAUD
External reset type: Contact closure
Software support: LabVIEW™ driver
Alarms
Number: 2
Settings: High and low set point, Inside/Outside, Audible
Actuators: Display annunciator, beeper, relays
Relays
Number: 3
Contacts: Normally open (NO), normally closed (NC), and common (C)
Contact rating: 30 VDC at 2 A
Operation: Follow high, low alarms with third relay indicating no alarm state — can be operated manually
Connector: Detachable terminal block
Monitor analog output
Scale: ±3 V = ±FS on Vs range
Accuracy: ±1% of reading ±10 mV, (DC to 10 kHz); ±5% of reading ±10 mv, (10 kHz to 50 kHz)
Minimum load resistance: 1 kΩ
Connector: Detachable terminal block
Corrected analog output
Scale: User selected
Range: ±10 V
Resolution: 0.3 mV
Accuracy: ±2.5 mV
Minimum load resistance: 1 kΩ
Connector: Detachable terminal block/p>
General
Ambient temperature: 15 ºC to 35 ºC at rated accuracy; 5 ºC to 40 ºC with reduced accuracy
Power requirement: 100, 120, 220, 240 VAC, +5%, -10%, 50 or 60 Hz, 20 VA
Size: 216 mm W × 89 mm H × 318 mm D (8.5 in × 3.5 in × 12.5 in), half rack
Weight: 3 kg (6.6 lb)
Approval: CE mark, RoHS