Thermometry |
Number of inputs |
4 (8 with scanner option) |
Input configuration |
Inputs can be configured from the front panel to accept any of the supported input types. Thermocouple and capacitance inputs require an optional input card that can be installed in the field. |
Supported option cards |
Thermocouple (3060), capacitance (3061), or scanner (3062) |
Option slots |
1 |
Isolation |
Sensor inputs optically isolated from other circuits but not from each other |
A/D resolution |
24-bit |
Input accuracy |
Sensor dependent, refer to Input Specifications table |
Measurement resolution |
Sensor dependent, refer to Input Specifications table |
Maximum update rate |
10 rdg/s on each input, 5 rdg/s when configured as 100 kΩ NTC RTD with reversal on, 2 rdg/s on each scanned input (scanner option only) |
Maximum update rate (scanner) |
The maximum update rate for a scanned input is 10 rdgs/s distributed among the enabled channels. Any channel configured as 100 kΩ RTD with reversal on changes the update rate for the channel to 5 rdgs/s.
Number of scanner channels enabled* |
Update rate |
1 |
10 rdg/s (100 ms/rdg) |
2 |
5 rdg/s (200 ms/rdg) |
3 |
3⅓ rdg/s (300 ms/rdg) |
4 |
2½ rdg/s (400 ms/rdg) |
5 |
2 rdg/s (500 ms/rdg) |
*No channels configured for 100 kΩ NTC RTD |
Autorange |
Automatically selects appropriate NTC RTD or PTC RTD range |
User curves |
Room for 39 200-point CalCurves™ or user curves |
SoftCal™ |
Improves accuracy of DT-470 diode to ±0.25 K from 30 K to 375 K; improves accuracy of platinum RTDs to ±0.25 K from 70 K to 325 K; stored as user curves |
Math |
Maximum and minimum |
Filter |
Averages 2 to 64 input readings |
|
Sensor Input Configuration |
|
Diode/RTD |
Thermocouple |
Measurement type |
4-lead differential |
2-lead differential, room temperature compensated |
Excitation |
Constant current with current reversal for RTDs |
NA |
SupportedSupported |
Diodes: Silicon, GaAlAs RTDs: 100 Ω Platinum, 1000 Ω Platinum, Germanium, Carbon-glass, Cernox®, and Rox™ |
Most thermocouple types |
Standard curves |
DT-470, DT-670, DT-500-D, DT-500-E1, PT-100, PT-1000, RX-102A, RX-202A |
Type E, Type K, Type T, AuFe 0.07% vs. Cr, AuFe 0.03% vs. Cr |
Input connector |
6-pin DIN |
Screw terminals in a ceramic isothermal block |
|
Control |
|
Control outputs |
4 |
|
|
Heater Outputs (Outputs 1 & 2) |
Control type |
Closed loop digital PID with manual heater output or open loop |
Update rate |
10/s |
Tuning |
Autotune (one loop at a time), PID, PID zones |
Control stability |
Sensor dependent, see Input Specifications table |
PID control settings |
Proportional (gain) |
0 to 1000 with 0.1 setting resolution |
Integral (reset) |
1 to 1000 (1000/s) with 0.1 setting resolution |
Derivative (rate) |
1 to 200% with 1% resolution |
Manual output |
0 to 100% with 0.01% setting resolution |
Zone control |
10 temperature zones with P, I, D, manual heater out, heater range, control channel, ramp rate |
Setpoint ramping |
0.1 K/min to 100 K/min |
|
Output 1 |
|
25 Ω setting |
50 Ω setting |
Type |
Variable DC current source |
D/A resolution |
16-bit |
Max power |
100 W |
50 W |
Max current |
2 A |
1 A |
Voltage complaince |
50 V |
50 V |
Heater load for max power |
25 Ω |
50 Ω |
Heater load range |
10 Ω to 100 Ω |
Ranges |
3 (decade steps in power) |
Heater noise |
0.12 µA RMS (dominated by line frequency and its harmonics) |
Grounding |
Output referenced to chassis ground |
Heater connector |
Dual banana |
Safety limits |
Curve temperature, power up heater off, short circuit protection |
|
Output 2 |
|
25 Ω setting |
50 Ω setting |
Type |
Variable DC current source |
D/A resolution |
16-bit |
Max power |
50 W |
50 W |
Max current |
1.41 A |
1 A |
Voltage compliance |
35.4 V |
50 V |
Heater load range |
10 Ω to 100 Ω |
Heater load for max power |
25 Ω |
50 Ω |
Ranges |
3 (decade steps in power) |
Heater noise |
0.12 µA RMS (dominated by line frequency and its harmonics) |
Grounding |
Output referenced to chassis ground |
Heater connector |
Dual banana |
Safety limits |
Curve temperature, power up heater off, short circuit protection |
|
|
|
|
Unpowered Outputs (Outputs 3 & 4) |
Control type |
Closed loop PID, PID zones, warm up heater mode, manual output, or monitor output |
Tuning |
Autotune (one loop at a time), PID, PID zones |
Control stability |
Sensor dependent, refer to Input Specifications table |
PID control settings |
|
Proportional (gain) |
0 to 1000 with 0.1 setting resolution |
Integral (reset) |
0 to 1000 (1000/s) with 0.1 setting resolution |
Derivative (rate) |
0 to 200% with 1% resolution |
Manual output |
0 to 100% with 0.01% setting resolution |
Zone control |
10 temperature zones with P, I, D, manual heater out, heater range, control channel, ramp rate |
Setpoint ramping |
0.1 K/min to 100 K/min |
Warm up heater mode settings
|
Warm up on percentage |
0 to 100% with 1% resolution |
Warm up mode
|
Continuous control or auto-off |
Analog output settings
|
Scale |
User selected |
Data source |
Temperature or sensor units |
Settings
|
Input, source, top of scale, bottom of scale, or manual |
Type
|
Variable DC voltage source |
Update rate
|
10/s |
Range
|
±10 V |
Resolution
|
16-bit, 0.3 mV |
Accuracy
|
±2.5 mV |
Noise
|
0.3 mV RMS |
Minimum load resistance
|
1 kΩ (short-circuit protected) |
Connector |
Detachable terminal block |
|
Front Panel |
|
Display
|
8-line by 40-character (240 × 64 pixel) graphic LCD display module with LED backlight |
Number of reading displays
|
1 to 8 |
Display units
|
K, °C, V, mV, Ω |
Reading source
|
Temperature, sensor units, max, and min |
Display update rate
|
2 rdg/s |
Temperature display resolution
|
0.0001° from 0° to 99.9999°, 0.001° from 100° to 999.999°, 0.01° above 1000° |
Sensor units display resolution
|
Sensor dependent, to 6 digits |
Other displays
|
Input name, setpoint, heater range, heater output, and PID |
Setpoint setting resolution
|
Same as display resolution (actual resolution is sensor dependent) |
Heater output display
|
Numeric display in percent of full scale for power or current |
Heater output resolution
|
0.01% |
Display annunciators
|
Control input, alarm, tuning |
LED annunciators
|
Remote, Ethernet status, alarm, control outputs |
Keypad |
27-key silicone elastomer keypad |
Front panel features |
Front panel curve entry, display contrast control, and keypad lock-out |
|
Interface |
|
IEEE-488.2
|
Capabilities |
SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT0, C0, E1 |
Reading rate |
To 10 rdg/s on each input |
Software Support
|
LabVIEW™ driver (contact Lake Shore for availability) |
USB
|
Function |
Emulates a standard RS-232 serial port |
Baud rate |
57,600 |
Connector |
B-type USB connector |
Reading rate |
To 10 rdg/s on each input |
Software Support |
LabVIEW™ driver (contact Lake Shore for availability) |
Ethernet |
|
Function |
TCP/IP, web interface, curve handler, configuration backup, chart recorder |
Connector |
RJ-45 |
Reading rate |
To 10 rdg/s on each input |
Software Support
|
LabVIEW™ driver (contact Lake Shore for availability) |
Alarms |
|
Number |
4 (8 with scanner option), high and low for each input |
Data source |
Temperature or sensor units |
Settings |
Source, high setpoint, low setpoint, deadband, latching or non- latching, audible on/off, and visible on/off |
Actuators
|
Display annunciator, beeper, and relays |
Relays |
|
Number |
2 |
Contacts |
Normally open (NO), normally closed (NC), and common (C) |
Contact Rating |
30 VDC at 3 A |
Operation |
Activate relays on high, low, or both alarms for any input, or manual mode |
Connector |
Detachable terminal block |
|
General |
|
Ambient temperature |
15 °C to 35 °C at rated accuracy; 5 °C to 40 °C at reduced accuracy |
Power requirement |
100, 120, 220, 240 VAC, ±10%, 50 or 60 Hz, 250 VA |
Size |
435 mm W × 89 mm H × 368 mm D (17 in × 3.5 in × 14.5 in), full rack |
Weight |
7.6 kg (16.8 lb) |
Approval |
CE mark, RoHS |
|
|