Specifications
Standard curve1 102 and 202: 0.05 K to 40 K; 103: 1.4 K to 40 K
Recommended excitation2 RX-102 and RX-202 : 20 µV (0.05 K to 0.1 K); 63 µV (0.1 K to
1.2 K); 10 mV or less for T > 1 K. RX-103: 10 mV or less for T > 1 K.
Dissipation at recommended excitation
102 and 202: 7.5 × 10–8 W at 4.2 K; 103: 3.2 ×
10–9 W at 1.4 K, 5.5 × 10–9 W at 4.2 K, 9.6 ×
10–9 W at 77 K
Thermal response time 0.5 s at 4.2 K, 2.5 s at 77 K
Radiation effects Recommended – Click here for more information
Magnetic field Recommended – Click here for more information
Reproducibility4 ±15 mK
1 102B does not follow a standard curve
2 Recommended excitation for T < 1 K based on Lake Shore calibration procedures using an AC resistance bridge – for more information refer to Appendix D and Appendix E
3 102B not recommended for use in magnetic fields
4 Short-term reproducibility data is obtained by subjecting sensor to repeated thermal shocks from 305 K to 4.2 K |