Contact: info@lakeshore.com Lakeshore.com Search www.lakeshore.com Contact Us Ordering / Pricing Products Home
Add To Cart
Home > Products > Cryogenic Temperature Sensors > Rhodium-Iron RTDs
 
  Rhodium-Iron RTDs
 
 

Specifications

Standard curve Not applicable

Recommended excitation Unpackaged chip and RF-100: 0.1 mA; RF-800: 1mA

Dissipation at recommended excitation
10 μW at 4.2 K, 250 μW at 273 K

Thermal response time Unpackaged chip: 2 ms at 4.2 K, 12 ms at 77 K, 35 ms at 273 K;
RF-100: 0.8 s at 4.2 K, 3.6 s at 77 K,
15 s at 273 K; RF-800: 10 s at 273 K

Use in radiation Recommended for use in ionizing radiation environments - Click here for more information

Use in magnetic field Not recommended for use in magnetic fields below 77 K – Click here for more information

Reproducibility1 RF-100: ±10 mK at 4.2 K;
RF-800: ±5 mK at 4.2 K

1 Short-term reproducibility data is obtained by subjecting sensor to repeated thermal shocks from 305 K to 4.2 K

 
Range of Use
 
Minimum
Limit
Maximum Limit
Unpackaged Chip
1.4 K
325 K
RF-100-AA
1.4 K
325 K
RF-800-4
0.65 K
500 K2
2 Usable to 800 K, but large and erratic temperature shifts can occur at lower temperature without proper thermal conditioning
Calibrated Accuracy: Unpackaged chip & RF-100
 

Typical sensor accuracy3

Long-term
stability4
1.4 K
±11 mK
±20 mK
4.2 K
±11 mK
±20 mK
10 K
±12 mK
±20 mK
77 K
±15 mK
±20 mK
305 K
±25 mK
±20 mK
Calibrated Accuracy: RF-800
 

Typical sensor accuracy3

Long-term
stability4
1.4 K
±7 mK
±10 mK
4.2 K
±7 mK
±10 mK
10 K
±8 mK
±10 mK
77 K
±13 mK
±10 mK
305 K
±23 mK
±10 mK
400 K
±41 mK
500 K
±42 mK
3 [(Calibration uncertainty)2 + (reproducibility)2] 0.5 for more information see Appendices B, D, and E
4 Long-term stability data is obtained by subjecting sensor to 200 thermal shocks from 305 K to 77 K
 
 
Temperature Response Data Table (typical)
 
RF-100
RF-800-4
R(W)
dR/dT
(W/K)
(T/R)·
(dR/dT)
R(W)
dR/dT
(W/K)
(T/R)·
(dR/dT)
1.4 K
6.892
0.489
0.099
1.5204
0.178
0.16
4.2 K
8.2053
0.418
0.21
1.9577
0.135
0.29
20 K
11.858
0.137
0.23
3.1632
0.0461
0.29
77 K
25.298
0.368
1.1
6.8341
0.0959
1.1
150 K
54.292
0.396
1.1
14.463
0.105
1.1
300 K
111.19
0.373
1.0
29.697
0.101
1.0
400 K
39.824
0.103
1.0
Click here for expanded response table
 
Physical Specifications
 
Size
Mass
Lead type
Internal
atmosphere
Sensor
materials
Bare Chip
(BR, BG
& BC)

1.3 mm wide ×
3.8 mm long ×
0.38 mm high
7 mg
BR: none;
BG: 4 gold;
BC: 4 copper;
– no polarity
NA
Rhodium-iron film chip with back side metallized with Mo/Au – if soldered attachment is desired, only indium solder should be used
RF-100-AA
3 mm diam ×
8.5 mm long
416 mg
4 phosphor bronze with heavy build polyimide attached with epoxy strain relief at sensor – color coded polarity
Helium 4 (4He) standard
Rhodium-iron chip is mounted strain-free in a cylindrical gold plated copper can
RF-800-4
0.51 mm ×
9 mm long
735 mg
4 platinum wire
Solid
Alumina and glass cylindrical case – rhodium-iron alloy wire encapsulated in ceramic

Typical Magnetic Field-Dependent Temperature Errors5 ΔT/T (%) at B (magnetic induction)

Package Parallel to Field B
T(K)
2.5 T
8 T
14 T
19 T
4.2
11
40
20
4
40
1.5
12
30
47
66
0.3
2.5
6
9
87
0.2
1.5
4
6
110
0.1
0.9
2.4
190
0.03
0.3
0.9
300
-0.01
0.1
0.4
5 Not recommended for use in magnetic fields below 77 K
Click to Enlarge
 
Click to Enlarge
 
 
 
 
 
 Home     |     Products     |     Ordering/Pricing     |     Search     |     Site Map
 
Privacy Policy | ©2007 Lake Shore Cryotronics, Inc.
 
Product Overview Ordering Information Downloads