Specifications
Standard curve IEC 751
Recommended excitation 1 mA
Dissipation at recommended excitation
100 µW at 273 K
Thermal response time
PT-102 & PT-103: 1.75 s at 77 K, 12.5 s at 273 K; PT-111: 2.5 s at 77 K, 20 s at 273 K
Use in radiation Recommended for use in ionizing radiation environments – Click here for more information
Use in magnetic field Because of their relatively low magnetic field dependence above 30 K, platinum sensors are useful as control elements in magnetic field applications when some error can be tolerated – Click here for more information
Reproducibility4 ±5 mK at 77 K
1 Short-term reproducibility data is obtained by subjecting sensor to repeated thermal shocks
from 305 K to 77 K |