Contact: info@lakeshore.com Lakeshore.com Search www.lakeshore.com Contact Us Ordering / Pricing Products Home
Add To Cart View Cart
Home > Products > Temperature Instruments > Monitors > Model 211
 
 Model 211 Temperature Monitor
 
 
 
Sensor Selection
 
Typical Sensor Performance - click here for sample calculations of typical sensor performance
 
 
 
Example
Lake Shore Sensor
Temp
Nominal
Resistance/
Voltage
Typical
Sensor Sensitivity6
Measurement
Resolution:
Temperature
Equivalents
Electronic Accuracy: Temperature Equivalents
Temperature Accuracy including Electronic Accuracy, CalCurve™, and Calibrated Sensor
Silicon Diode
DT-670-SD
with 1.4H
calibration
1.4 K
1.644 V
-12.49 mV/K
1.6 mK
±26 mK
±38 mK
77 K
1.028 V
-1.73 mV/K
11.6 mK
±152 mK
±174 mK
300 K
0.5597 V
-2.3 mV/K
8.7 mK
±94 mK
±126 mK
500 K
0.0907 V
-2.12 mV/K
9.4 mK
±80 mK
±130 mK
Silicon Diode
DT-470-SD-13
with 1.4H calibration
1.4 K
1.6981 V
-13.1 mV/K
1.5 mK
±26 mK
±38 mK
77 K
1.0203 V
-1.92 mV/K
10.5 mK
±137 mK
±159 mK
300 K
0.5189 V
-2.4 mV/K
8.4 mK
±88 mK
±120 mK
475 K
0.0906 V
-2.22 mV/K
9.1 mK
±77 mK
±127 mK
GaAlAs Diode
TG-120-SD
with 1.4H calibration
1.4 K
5.391 V
-97.5 mV/K
0.2 mK
±13 mK
±25 mK
77 K
1.422 V
-1.24 mV/K
16.2 mK
±359 mK
±381 mK
300 K
0.8978 V
-2.85 mV/K
7 mK
±120 mK
±152 mK
475 K
0.3778 V
-3.15 mV/K
6.4 mK
±75 mK
±125 mK
100 W Platinum RTD
500 W Ful Scale
PT-103
with 1.4J calibration
30 K
3.66 W
0.19 W/K
10.5 mK
±25 mK
±35 mK
77 K
20.38 W
0.42 W/K
4.8 mK
±20 mK
±32 mK
300 K
110.35 W
0.39 W/K
5.2 mK
±68 mK
±91 mK
500 K
185.668 W
0.378 W/K
5.3 mK
±109 mK
±155 mK
Cernox™
CX-1050-SD-HT7 with 4M
calibration
4.2 K
3507.2 W
-1120.8 W/K
45 µK
±1.4 mK
±6.4 mK
77 K
205.67 W
-2.4116 W/K
20.8 mK
±75.6 mK
±91.6 mK
300 K
59.467 W
-0.1727 W/K
290 mK
±717 mK
±757 mK
420 K
45.03 W
-0.0829 W/K
604 mK
±1.43 K
±1.5 K
Germanium
GR-200A-1000 with 1.4D calibration
2 K
6674 W
-9930 W/K
5 µK
±0.3 mK
±4.3 mK
4.2 K
1054 W
-526 W/K
95 µK
±10 mK
±14 mK
10 K
170.9 W
-38.4 W/K
1.3 mK
±4.4 mK
±9.4 mK
100 K
2.257 W
-0.018 W/K
2.78 K
±5.61 K
±5.77 K
Carbon-Glass
CGR-1-2000
with 4L
calibration
4.2 K
2260 W
-2060 W/K
25 µK
±0.5 mK
±4.5 mK
77 K
21.65 W
-0.157 W/K
319 mK
±692 mK
±717 mK
300 K
11.99 W
-0.015 W/K
3.33 K
±7 K
±7.1 K

6 Typical sensor sensitivities were taken from representative calibrations for the sensor listed
7 Non-HT version maximum temperature: 325 K

 
Downloads Ordering Information Technical Specifications
 