| Thermometry |
| Number of inputs |
1 |
| Input configuration |
Inputs can be configured from the front panel to accept any of the supported input types |
| Isolation |
Measurement is not isolated from chassis ground |
| A/D resolution |
24-bit |
| Input accuracy |
Sensor dependent – refer to Input Specifications table |
| Measurement resolution |
Sensor dependent – refer to Input Specifications table |
| Maximum update rate |
7 rdg/s |
| User curve |
One 200-point CalCurve™ or user curve in non-volatile memory |
 |
 |
|
| |
| Sensor Input Configuration |
| |
Diode/RTD |
| Measurement type |
4-lead differential |
| Excitation |
Constant current |
| Supported sensors |
Diodes: Silicon, GaAlAs
RTDs: 100 W Platinum, 1000 W Platinum,
Carbon-Glass, Cernox™ , and Rox™ |
| Standard curves |
DT-470, DT-670, CTI-C, PT-100, and PT-1000 |
| Input connector |
Shared 25-pin D-sub |
 |
 |
|
| Front Panel |
|
| Display |
5-digit LED |
Number of
reading displays |
1 |
| Display Units |
K, ºC, ºF, V, and W |
| Reading source |
Temperature and sensor units |
| Display update rate |
2 rdg/s |
| Temp display resolution |
0.001º from 0º to 99.999º, 0.01º from 100º to 999.99º, 0.1º above 1000º |
| Sensor units display resolution |
Sensor dependent, to 5 digits |
| Display annuciators |
K, °C, °F, V/W |
| Keypad |
4 full travel keys, numeric and specific functions |
| Front panel features |
Display brightness control, keypad lock-out |
 |
 |
|
| Interface |
|
| Serial interface |
| Electrical format |
RS-232C |
| Max baud rate |
9600 baud |
| Connector |
9-pin D-sub |
| Reading rate |
Up to 7 rdg/s |
| Alarms |
| Number |
2, high and low |
| Data source |
Temperature |
| Settings |
High setpoint, Low setpoint, Dead band, Latching or Non-latching |
| Actuators |
Display message, relays |
| Relays |
| Number |
2 |
| Contacts |
Normally open (NO), normally closed (NC), and common (C) |
| Contact rating |
30 VDC at 1 A |
| Operation |
Activate relays on high or low input alarm or manual |
| Connector |
Shared 25-pin D-sub |
| Analog output |
| Isolation |
Output is not isolated from chassis ground |
| Update rate |
7 rdg/s |
| Data source |
Temperature |
| |
Voltage |
Current |
| Range |
0 V to 10 V |
4 mA to 20 mA |
| Accuracy |
±1.25 mV |
±2.5 µA |
| Resolution |
0.3 mV |
0.6 µV |
| Min load resistance |
500 W |
NA |
| Compliance voltage |
NA |
10 V |
| Load regulation |
NA |
±0.02% of reading
0 to 500 W |
|
|
 |
 |
 |
|
|
| |
Temperature |
Sensor units
(fixed by type) |
| Scales: |
0 K to 20 K |
Diodes: 1 V = 1 V |
| |
0 K to 100 K |
100 W Platinum:
1 V = 100 W |
| |
0 K to 200 K |
1000 W Platinum:
1 V = 1000 W |
| |
0 K to 325 K |
NTC Resistor:
1 V = 1000 W |
| |
0 K to 475 K |
|
| |
0 K to 1000 K |
|
 |
 |
 |
|
| Settings |
Voltage or current, scale |
| Connector |
Shared 25-pin D-sub |
 |
 |
|
| |
| .......................................................................................... |