 Model 336 Cryogenic Temperature Controller

 
 
 
Sensor Temperature Range (sensors sold separately)
 
Silicon diodes are the best choice for general cryogenic use from 1.4 K to above room temperature. Silicon Diodes are economical to use because they follow a standard curve and are interchangeable in many applications. They are not suitable for use in ionizing radiation or magnetic fields.
 
Cernox™ thin-film RTDs offer high sensitivity and low magnetic field-induced errors over the 0.3 K to 420 K temperature range. Cernox sensors require calibration.
 

Platinum RTDs offer high uniform sensitivity from 30 K to over 800 K. With excellent reproducibility, they are useful as thermometry standards. They follow a standard curve above 70 K and are interchangeable in many applications.

 

1 Non-HT version maximum temperature: 325 K
2 Low temperature limited by input resistance range

3 Low temperature specified with self-heating error:
5 mK

 
   
Model
Useful Range
Magnetic Field Use
Diodes
Silicon Diode
DT-670-SD
1.4 K to 500 K
T 60 K & B 3 T
Silicon Diode
DT-670E-BR
30 K to 500 K
T 60 K & B 3 T
Silicon Diode
DT-414
1.4 K to 375 K
T 60 K & B 3 T
Silicon Diode
DT-421
1.4 K to 325 K
T 60 K & B 3 T
Silicon Diode
DT-470-SD
1.4 K to 500 K
T 60 K & B 3 T
Silicon Diode
DT-471-SD
10 K to 500 K
T 60 K & B 3 T
GaAlAs Diode
TG-120-P
1.4 K to 325 K
T > 4.2 K & B 5 T
GaAlAs Diode
TG-120-PL
1.4 K to 325 K
T > 4.2 K & B 5 T
GaAlAs Diode
TG-120-SD
1.4 K to 500 K
T > 4.2 K & B 5 T
Positive Temperature
Coefficient RTDs
100 W Platinum
PT-102/3
14 K to 873 K
T > 40 K & B 2.5 T
100 W Platinum
PT-111
14 K to 673 K
T > 40 K & B 2.5 T
Rhodium-Iron
RF-800-4
1.4 K to 500 K
T > 77 K & B 8 T
Rhodium-Iron
RF-100T/U
1.4 K to 325 K
T > 77 K & B 8 T
Negative Temperature
Coefficient RTDs
Cernox™
CX-1010
0.3 K to 325 K1
T > 2 K & B 19 T
Cernox™
CX-1030-HT
0.3 K to 420 K1,3
T > 2 K & B 19 T
Cernox™
CX-1050-HT
1.4 K to 420 K1
T > 2 K & B 19 T
Cernox™
CX-1070-HT
4 K to 420 K1
T > 2 K & B 19 T
Cernox™
CX-1080-HT
20 K to 420 K1
T > 2 K & B 19 T
Germanium
GR-200A-100
0.3 K to 100 K
Not Recommended
Germanium
GR-200A-250
0.5 K to 100 K
Not Recommended
Germanium
GR-200A/B-500
1.4 K to 100 K
Not Recommended
Germanium
GR-200A/B-1000
1.4 K to 100 K
Not Recommended
Germanium
GR-200A/B-1500
1.4 K to 100 K
Not Recommended
Germanium
GR-200A/B-2500
1.4 K to 100 K
Not Recommended
Carbon-Glass
CGR-1-500
1.4 K to 325 K
T > 2 K & B 19 T
Carbon-Glass
CGR-1-1000
1.7 K to 325 K2
T > 2 K & B 19 T
Carbon-Glass
CGR-1-2000
2 K to 325 K2
T > 2 K & B 19 T
Rox™
RX-102
0.3 K to 40 K3
T > 2 K & B 10 T
Rox™
RX-103
1.4 K to 40 K
T > 2 K & B 10 T
Rox™
RX-202
0.3 K to 40 K3
T > 2 K & B 10 T
Thermocouples
Type K
9006-006
3.2 K to 1505 K
Not Recommended
Type E
9006-004
3.2 K to 934 K
Not Recommended
Chromel-AuFe 0.07%
9006-002
1.2 K to 610 K
Not Recommended
 
Downloads for Model 336 Cryogenic Temperature Controller Ordering Information for Model 336 Cryogenic Temperature Controller Technical Specifications for Model 336 Cryogenic Temperature Controller
 