Add Model 332 Cryogenic Temperature Controller To Cart View Cart
Home > Products > Cryogenic Temperature Instruments > Cryogenic Temperature Controllers > Model 332
 

 Model 332 Cryogenic Temperature Controller

 
 
 Input Specifications
 
Sensor
Temperature
Coefficient
Input
Range
Excitation
Current
Display
Resolution
Measurement
Resolution
Electronic
Accuracy
Electronic
Control
Stability12
Diode
negative
0 V to 2.5 V
10 µA
±0.05%14, 15
100 µV
10 µV
±80 µV ±0.005% of rdg
±20 µV
negative
0 V to 7.5 V
10 µA
±0.05%14, 15
100 µV
20 µV
±80 µV ±0.005% of rdg
±40 µV
PTC RTD
positive
0 W to 250 W
1 mA16
10 mW
2 mW
±0.004W ±0.01% of rdg
±4 mW
positive
0 W to 500 W
1 mA16
10 mW
2 mW
±0.004W ±0.01% of rdg
±4 mW
positive
0 W to 5000 W
1 mA16
100 mW
20 mW
±0.004W ±0.02% of rdg
±40 mW
NTC RTD
negative
0 W to 75 W
1 mA16
1 mW
0.3 mW +
0.000% of rdg
±0.001 W ±0.04% of rdg
±0.6 mW
negative
0 W to 750 W
100 µA16
10 mW
3 mW +
0.001% of rdg
±0.01 W ±0.04% of rdg
±6 W
±0.002%
of rdg
negative
0 W to 7500 W
10 µA16
100 mW
20 mW +
0.001% of rdg
±0.1 W ±0.04% of rdg
±40 mW
±0.002%
of rdg
negative
0 W to 75000 W
1 µA16
1 W
0.15 W +
0.003% of rdg
±1.0 W ±0.04% of rdg
±0.3 W
±0.006%
of rdg
Thermocouple
positive
±25 mV
NA
1 µV
0.4 µV
±1 µV ±0.05% of rdg17
0.8 µV
positive
±50 mV
NA
1 µV
0.4 µV
±1 µV ±0.05% of rdg17
0.8 µV
13 Control stability of the electronics only, in an ideal thermal system
14 Current source error has negligible effect on measurement accuracy
15 Diode input excitation current can be set to 1 mA – refer to the Model 331 user manual for details
16 Current source error is removed during calibration
17 Accuracy specification does not include errors from room temperature compensation
 
Thermometry
Number of inputs 2
Input configuration Each input is factory configured as either diode/RTD or thermocouples
Isolation Sensor inputs optically isolated from other circuits but not from each other
A/D resolution 24-bit
Input accuracy Sensor dependent – refer to Input Specifications table
Measurement resolution Sensor dependent – refer to Input Specifications table
Maximum update rate 10 readings per s on each input with the following exceptions: 5 readings per s when configured as 75 kW NTC RTD with reversal on, 5 readings per s on input A when configured as thermocouple
Autorange Automatically selects appropriate NTC RTD range
User curves Room for 20 200-point CalCurves™ or user curves
SoftCal™ Improves accuracy of DT-470 diode to ±0.25 K from 30 K to 375 K; improves accuracy of platinum RTDs to ±0.25 K from 70 K to 325 K; stored as user curves
Math Maximum and minimum, and linear equation
(Mx + B) or M(x + B)
Filter Averages 2 to 64 input readings
 
Sensor Input Configuration
 
Diode/RTD
Thermocouple
Measurement type
4-lead differential
2-lead, room temperature compensated
Excitation
Constant current with current reversal for RTDs
NA
Supported
sensors
Diodes: Silicon, GaAlAs
RTDs: 100 W Platinum, 1000 W Platinum, Germanium, Carbon-Glass, Cernox™, and Rox™
Most thermocouple types
Standard
curves
DT-470, DT-500D, DT-670, PT-100, PT-1000, RX-102A,
RX-202A
Type E, Type K, Type T AuFe 0.07% vs. Cr,
AuFe 0.03% vs Cr,
Input
connector
6-pin DIN
Ceramic isothermal block
Control  
Control loops 2
Control type Closed-loop digital PID with manual heater output or open loop
Tuning Autotune (one loop at a time), manual PID, zones
Control stability Sensor dependent – to 2× measurement resolution (in an ideal thermal system)
PID control settings
    Proportional (gain) 0 to 1000 with 0.1 setting resolution
    Integral (reset) 1 to 1000 (1000 per s) with 0.1 setting resolution
    Derivative (rate) 1% to 200% with 1% setting resolution
    Manual output 0% to 100% with 0.001% setting resolution
Zone control 10 temperature zones with P, I, D, manual heater out, and heater range
Setpoint ramping 0.1 K per min to 100 K per min
Safety limits Curve temperature, power up heater off, and short-circuit protection
 
Heater Output
 
Loop 1
Loop 2
Heater output type
Variable DC current source
Variable DC voltage source
Heater output D/A resolution
18-bit
16-bit
Max heater power
50 W
10 W
Max heater output current
1 A
1 A
Heater output complaince
50 V
10 V
Heater source impedance
N/A
0.1 W maximum
Heater output ranges
3 decade steps in power
1
Heater load type
Resistive
Resistive
Heater load range
10 W to 100 W recommended
10 W minimum
Heater load for max power
50 W
10 W
Heater noise (<1 kHz) RMS
50 µV + 0.017% of output voltage
<0.3 mV
Isolation
Optical isolation between output and other circuits
None
Heater connector
Dual banana
Detachable terminal block
 
Loop 1 Full Scale Heater Power at Typical Resistance
Heater Resistance
Heater Range
Heater Power
10 W
Low
100 mW
Med
1 W
High
10 W
25 W
Low
250 mW
Med
2.5 W
High
25 W
50 W
Low
500 mW
Med
5 W
High
50 W
...........................................................................................
You may also be interested in...
Sensors
Measure temperatures from <20 mK to over 1,500 K. A wide selection of diodes, RTDs, thermocouples, and mounting packages are available. Click here for more information on cryogenic temperature sensors.
Cryogenic Accessories
Cryogenic and temperature measurement accessories include cable, wire, solder, heaters, insulating and conductive materials, and more. Click here for more information on cryogenic accessories.
Front Panel  
Display 2 line by 20 character, 9 mm character height, vacuum fluorescent display
Number of reading displays 1 to 4
Display Units K, ºC, V, mV and W
Reading source Temperature, sensor units, max, min, and linear equation
Display update rate All readings twice per s
Temp display resolution 0.001º from 0º to 99.999º, 0.01º from 100º to 999.99º, 0.1º above 1000º
Sensor units display resolution Sensor dependent to 5 digits
Other displays Setpoint, heater range, and heater output
(user selected)
Setpoint setting resolution Same as display resolution
(actual resolution is sensor dependent)
Heater output display Numerical or graphical display in percent of full scale for power or current
Heater output resolution 1% numeric or 2% graphical
Display annunciators Control input, remote, alarm, tuning, ramp, max, min, and linear
Keypad 20 full-travel keys, numeric and specific functions
Front panel features Front panel curve entry, display brightness control, and keypad lock-out
Interfaces  
IEEE-488.2 interface
    Features SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT0, C0, E1
    Reading rate To 10 readings per s on each point
    Software Support LabVIEW™ driver
Serial interface
    Electrical format RS-232C
    Max baud rate 9600 baud
    Connector 9-pin D-sub
    Reading rate To 10 readings per s on each input (at 9600 baud)
Special interface features Model 330 command emulation mode
Alarms
    Number 4: high and low for each input
    Data source Temperature, sensor units, and linear equation
    Settings Source, high setpoint, low setpoint, deadband, latching or non-latching, and audible on/off
    Actuators Display, annunciator, beeper, and relays
Relays  
    Number 2
    Contacts Normally open (NO), normally closed (NC), and (C) common
    Contact Rating 30 VDC at 5 A
    Operation Activate relays on high, low or both alarms for either input or manual
    Connector Detachable terminal block
Analog voltage outputs (when not used as control loop 2 output)
    Scale User selected
    Update rate 10 readings per s
    Data source Temperature, sensor units, and linear equation
    Settings Input, source, top of scale, bottom of scale, or manual
    Range ±10 V
    Resolution 0.3 mV
    Accuracy ±2.5 mV
    Max output power 1 W (jumper selected)
    Min load resistance 100 W (short-circuit protected)
    Source impedance 0.01 W
General  
Ambient temp range

15 °C to 35°C at rated accuracy, 10 ºC to 40 ºC at reduced accuracy

Power requirements 100, 120, 220, 240 VAC (+6%, -10%), 50 or 60 Hz, 150 VA
Size 216 mm W × 89 mm H × 368 mm D
(8.5 in × 3.5 in × 14.5 in), half rack
Weight 4.8 kg (10.5 lb)
Approval CE mark
 
 
Sensor Inputs
Temperature Control
Interface
Configurable Display
Sensor Selection: Sensor Temperature Range
Sensor Selection: Typical Sensor Performance
 
 Home     |     Products     |     Ordering/Pricing     |     Search     |     Site Map
 