| Thermometry |
| Number of inputs |
2 |
| Input configuration |
Each input is factory configured as either diode/RTD or thermocouples |
| Isolation |
Sensor inputs optically isolated from other circuits but not from each other |
| A/D resolution |
24-bit |
| Input accuracy |
Sensor dependent – refer to Input Specifications table |
| Measurement resolution |
Sensor dependent – refer to Input Specifications table |
| Maximum update rate |
10 readings per s on each input with the following exceptions: 5 readings per s when configured as 75 kW NTC RTD with reversal on, 5 readings per s on input A when configured as thermocouple |
| Autorange |
Automatically selects appropriate NTC RTD range |
| User curves |
Room for 20 200-point CalCurves™ or user curves |
| SoftCal™ |
Improves accuracy of DT-470 diode to ±0.25 K from 30 K to 375 K; improves accuracy of platinum RTDs to ±0.25 K from 70 K to 325 K; stored as user curves |
| Math |
Maximum and minimum, and linear equation
(Mx + B) or M(x + B) |
| Filter |
Averages 2 to 64 input readings |
 |
 |
|
| |
| Sensor Input Configuration |
| |
Diode/RTD |
Thermocouple |
| Measurement type |
4-lead
differential |
2-lead, room temperature compensated |
| Excitation |
Constant current with current reversal for RTDs |
NA |
Supported
sensors |
Diodes: Silicon, GaAlAs
RTDs: 100 W Platinum, 1000 W Platinum, Germanium, Carbon-Glass, Cernox™, and Rox™ |
Most thermocouple types |
Standard
curves |
DT-470, DT-500D, DT-670, PT-100, PT-1000, RX-102A,
RX-202A |
Type E, Type K, Type T AuFe 0.07% vs. Cr,
AuFe 0.03% vs Cr, |
Input
connector |
6-pin DIN |
Ceramic isothermal block |
 |
 |
 |
|
| Control |
|
| Control loops |
2 |
| Control type |
Closed-loop digital PID with manual heater output or open loop |
| Tuning |
Autotune (one loop at a time), manual PID, zones |
| Control stability |
Sensor dependent – to 2× measurement resolution (in an ideal thermal system) |
| PID control settings |
| Proportional (gain) |
0 to 1000 with 0.1 setting resolution |
| Integral (reset) |
1 to 1000 (1000 per s) with 0.1 setting resolution |
| Derivative (rate) |
1% to 200% with 1% setting resolution |
| Manual output |
0% to 100% with 0.001% setting resolution |
| Zone control |
10 temperature zones with P, I, D, manual heater out, and heater range |
| Setpoint ramping |
0.1 K per min to 100 K per min |
| Safety limits |
Curve temperature, power up heater off, and short-circuit protection |
 |
 |
|
| |
| Heater Output |
| |
Loop 1 |
Loop 2 |
| Heater output type |
Variable DC current source |
Variable DC voltage source |
| Heater output D/A resolution |
18-bit |
16-bit |
| Max heater power |
50 W |
10 W |
| Max heater output current |
1 A |
1 A |
| Heater output complaince |
50 V |
10 V |
| Heater source impedance |
N/A |
0.1 W maximum |
| Heater output ranges |
3 decade steps in power |
1 |
| Heater load type |
Resistive |
Resistive |
| Heater load range |
10 W to 100 W recommended |
10 W minimum |
| Heater load for max power |
50 W |
10 W |
| Heater noise (<1 kHz) RMS |
50 µV + 0.017% of output voltage |
<0.3 mV |
| Isolation |
Optical isolation between output and other circuits |
None |
| Heater connector |
Dual banana |
Detachable terminal block |
 |
 |
 |
|
| |
| Loop 1 Full Scale Heater Power at Typical Resistance |
Heater
Resistance |
Heater
Range |
Heater Power |
10 W |
Low |
100 mW |
Med |
1 W |
High |
10 W |
25 W |
Low |
250 mW |
Med |
2.5 W |
High |
25 W |
50 W |
Low |
500 mW |
Med |
5 W |
High |
50 W |
 |
 |
 |
|
| ........................................................................................... |
| You may also be interested in... |
|
|
| Front Panel |
|
| Display |
2 line by 20 character, 9 mm character height, vacuum fluorescent display |
| Number of reading displays |
1 to 4 |
| Display Units |
K, ºC, V, mV and W |
| Reading source |
Temperature, sensor units, max, min, and linear equation |
| Display update rate |
All readings twice per s |
| Temp display resolution |
0.001º from 0º to 99.999º, 0.01º from 100º to 999.99º, 0.1º above 1000º |
| Sensor units display resolution |
Sensor dependent to 5 digits |
| Other displays |
Setpoint, heater range, and heater output
(user selected) |
| Setpoint setting resolution |
Same as display resolution
(actual resolution is sensor dependent) |
| Heater output display |
Numerical or graphical display in percent of full scale for power or current |
| Heater output resolution |
1% numeric or 2% graphical |
| Display annunciators |
Control input, remote, alarm, tuning, ramp, max, min, and linear |
| Keypad |
20 full-travel keys, numeric and specific functions |
| Front panel features |
Front panel curve entry, display brightness control, and keypad lock-out |
 |
 |
|
| Interfaces |
|
| IEEE-488.2 interface |
| Features |
SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT0, C0, E1 |
| Reading rate |
To 10 readings per s on each point |
| Software Support |
LabVIEW™ driver |
| Serial interface |
| Electrical format |
RS-232C |
| Max baud rate |
9600 baud |
| Connector |
9-pin D-sub |
| Reading rate |
To 10 readings per s on each input (at 9600 baud) |
| Special interface features |
Model 330 command emulation mode |
| Alarms |
|
| Number |
4: high and low for each input |
| Data source |
Temperature, sensor units, and linear equation |
| Settings |
Source, high setpoint, low setpoint, deadband, latching or non-latching, and audible on/off |
| Actuators |
Display, annunciator, beeper, and relays |
| Relays |
|
| Number |
2 |
| Contacts |
Normally open (NO), normally closed (NC), and (C) common |
| Contact Rating |
30 VDC at 5 A |
| Operation |
Activate relays on high, low or both alarms for either input or manual |
| Connector |
Detachable terminal block |
| Analog voltage outputs (when not used as control loop 2 output) |
| Scale |
User selected |
| Update rate |
10 readings per s |
| Data source |
Temperature, sensor units, and linear equation |
| Settings |
Input, source, top of scale, bottom of scale, or manual |
| Range |
±10 V |
| Resolution |
0.3 mV |
| Accuracy |
±2.5 mV |
| Max output power |
1 W (jumper selected) |
| Min load resistance |
100 W (short-circuit protected) |
| Source impedance |
0.01 W |
 |
 |
|
| General |
|
| Ambient temp range |
15 °C to 35°C at rated accuracy, 10 ºC to 40 ºC at reduced accuracy |
| Power requirements |
100, 120, 220, 240 VAC (+6%, -10%), 50 or 60 Hz, 150 VA |
| Size |
216 mm W × 89 mm H × 368 mm D
(8.5 in × 3.5 in × 14.5 in), half rack |
| Weight |
4.8 kg (10.5 lb) |
| Approval |
CE mark |
 |
 |
|
| |
| |
Sensor Inputs |
Temperature Control |
Interface |
Configurable Display |
Sensor Selection: Sensor Temperature Range |
Sensor Selection: Typical Sensor Performance |
|
|