Contact: info@lakeshore.com Lakeshore.com Search www.lakeshore.com Contact Us Ordering / Pricing Products Home
Add Model 321 Cryogenic Temperature Controller To Cart View Cart
Home > Products > Cryogenic Temperature Instruments > Cryogenic Temperature Controllers > Model 321
 

 Model 321 Cryogenic Temperature Controller

 
 
 
The Model 321 temperature controller is obsolete and has been replaced by the new Model 325 temperature controller.
 
Sensor Selection
 
Typical Sensor Performance - click here for sample calculations of typical sensor performance
 
 
Example Lake Shore Sensor
Temp
Nominal
Resistance/
Voltage
Typical
Sensor Sensitivity2
Measurement
Resolution:
Temperature
Equivalents
Electronic Accuracy: Temperature Equivalents
Temperature Accuracy including Electronic Accuracy, CalCurve™, and Calibrated Sensor
Electronic Control Stability3:
Temperature Equivalents
Model 321-01
Silicon Diode
DT-670-C0-13 with 1.4H calibration
1.4 K
1.644 V
-12.49 mV/K
3.3 mK
±42.4 mK
±54.4 mK
±6.6 mK
77 K
1.028 V
-1.73 mV/K
23.1 mK
±234 mK
±246 mK
±46.2 mK
300 K
0.5597 V
-2.3 mV/K
17.4 mK
±135 mK
±167 mK
±34.8 mK
500 K
0.0907 V
-2.12 mV/K
18.9 mK
±103 mK
±153 mK
±37.8 mK
Model 321-01
Silicon Diode
DT-470-SD-13 with 1.4H calibration
1.4 K
1.6981 V
-13.1 mV/K
3.1 mK
±41.2 mK
±53.2 mK
±6.2 mK
77 K
1.0203 V
-1.92 mV/K
20.8 mK
±210 mK
±232 mK
±41.6 mK
300 K
0.5189 V
-2.4 mV/K
16.7 mK
±127 mK
±159 mK
±33.4 mK
475 K
0.0906 V
-2.22 mV/K
18.0 mK
±98.3 mK
±148 mK
±36.0 mK
Model 321-02
100 W Platinum RTD
300 W Full Scale
PT-103
with 1.4J calibration
30 K
3.660 W
0.191 W/K
26.2 mK
±115 mK
±125 mK
±52.4 mK
77 K
20.38 W
0.423 W/K
11.9 mK
±72 mK
±84 mK
±23.8 mK
300 K
110.35 W
0.387 W/K
13 mK
±195 mK
±218 mK
±26 mK
500 K
185.668 W
0.378 W/K
14 mK
±299 mK
±345 mK
±28 mK
Model 321-044 Thermocouple
45 mV
Chromel vs
AuFe 0.07%
2 K
-5292 µV
10.1 µV/K
149 mK
±658 mK5
Calibration not available from
Lake Shore
±298 mK
300 K
597.44 µV
22.4 µV/K
67 mK
±192 mK5
±134 mK
600 K
7470.7 µV
23.4 µV/K
65 mK
±331 mK5
±130 mK

2 Typical sensor sensitivities were taken from representative calibrations for the sensor listed
3 Control stability of the electronics only, in an ideal thermal system
4 All thermocouple data is for uncompensated inputs
5 Accuracy specification does not include errors from room temperature compensation

   
 
Downloads for Model 321 Cryogenic Temperature Controller Ordering Information for Model 321 Cryogenic Temperature Controller Technical Specifications for Model 321 Cryogenic Temperature Controller
 
 Home     |     Products     |     Ordering/Pricing     |     Search     |     Site Map
 