Contact: info@lakeshore.com Lakeshore.com Search www.lakeshore.com Contact Us Ordering / Pricing Products Home
Desert Web Logo
Add Model HFTTP4 Probe Station To Cart View Cart
 
Home > Lake Shore Products > Cryogenic Probe Stations > Model HFTTP4 Cryogenic Probe Station Technical Specifications
 

 Model HFTTP4 Cryogenic Probe Station

 
Model HFTTP4 Probe Station Product Overview Request A Quote for Model HFTTP4 Probe Station Downloads for Model HFTTP4 Probe Station
 
Specifications
 
Magnetic Field
Maximum field
 10 k0e (1 T)
 
Temperature
Operating temperature range—all probe types
 2 K to 400 K
Temperature control (heaters)
     Cold head stage  50 W
     Magnet stage  100 W
     Two radiation shields  100 W and 50 W
     Probe arm
  Measurement only
 
Probe Arm Adjustments
Travel
     X axis  51 mm (2 in)
     Y axis
 25 mm (1 in)
     Z axis
 18 mm (0.7 in)
Translation resolution
     X axis  20 µm
     Y and Z axes
 10 µm
     Theta planarization*
 ±5°
*Included with microwave probes  
 
Frequency Range
ZN50 DC/RF probe frequency range
          Tungsten with cryogenic coax cable  0 to 50 MHz*
          Tungsten with semirigid coax cable  0 to 1 GHz*†
          Paliney 7 with cryogenic coax cable  0 to 50 MHz*
          Paliney 7 with semirigid coax cable  0 to 1 GHz*†
          BeCu with cryogenic coax cable  0 to 50 MHz*
          BeCu with semirigid coax cable  0 to 1 GHz*†
GSG microwave probe frequency range
          Low frequency with K connector  0 to 40 GHz*
          Mid frequency with 2.4 mm connector  0 to 50 GHz*
          High frequency with 1.8 mm connector  0 to 67 GHz*
Click here for probe & cable features
*Selectable equipment
S21 > -10 dB up to 1 Ghz, except for a (-40 dB) spike between 400 MHz and 800 MHz depending on probe model and placement; S11 < -3 dB up to 1 GHz
 
Optical
Optical viewport—located on top lids
 Ø54 mm (2.1 in) outer window and Ø51 mm (2 in) inner window
Outer, clear fused quartz  99% IR transmittance
Inner
 IR absorbing with narrow band visible light transmittance
Optical resolution—monoscope
7:1 zoom  9 µm
12.5:1 zoom  5 µm*
16:1 zoom
 4 µm*
*Selectable equipment
 
Sample Holder (Chuck)
Maximum sample size—overall  Up to Ø25 mm (1 in)
          SH-1.00-G, Grounded chuck  Up to Ø25 mm (1 in) and 400 K
          SH-1.00-I, Isolated chuck  Up to Ø25 mm (1 in) and 400 K*
          SH-1.00-C, Coaxial chuck  Up to Ø25 mm (1 in) and 400 K*
          SH-1.00-T, Triaxial chuck  Up to Ø25 mm (1 in) and 400 K*
Click here to select a sample chuck (holder)  
*Selectable equipment
 
Standard Equipment
Superconducting magnet  10 k0e (1 T), horizontal field, split pair
 
Superconducting magnet power supply  Lake Shore Model 625
     Output type  Bipolar, 4-quadrant, DC current source
     Current  ±60 A
     Voltage  ±5 V
 
Open cycle multi-stage refrigerator
 4.2 K to 400 K
     Cold head stage temperature sensor
 Lake Shore Model CX-1030-SD-HT-4M calibrated CernoxTM RTD
     Cold head stage heater
 50 W
     Magnet stage temperature sensor  Lake Shore Model CX-1030-SD-HT-4M calibrated CernoxTM RTD
     Magnet stage heater  100 W
 
Cooled radiation shield and cooled IR-absorbing window above the sample
     Two radiation shield temperature sensors  Lake Shore Model DT-670C-CU silicon diode
     Two radiation shield heaters  100 W and 50 W
     Removable top lid with viewport  Ø51 mm (2 in) window
 
Temprature controllers  One Lake Shore Model 340 with 3462 expansion card, one Model 332S, and  one Model 142 200 W (2-channels, 100 W each) power supply (independent  regulation of cold head  stage, magnet stage, radiation shield, and probe  arm temperature  monitoring)
  
Vacuum chamber

     Diameter   
 25.4 cm (10 in)
     Removable top lid with clear fused quartz viewport  Ø54 mm (2.1 in) window
     Probe ports  4 equally spaced surround the sample thermal radiation shield
  
Machined aluminum base plate  53.3 cm2 (21 in2)
 
Vibration isolation table  Low vibration, non-magnetic stand, PVC construction with sand weighted  legs and brass fittings
 
Four XYZ precision micro-manipulated probing stages
 
Probe arms, thermal radiation shields, stainless steel welded bellows, and feedthrough ports
 
One probe mount includes a platinum temperature sensor installed and wired  to a 6-pin feedthrough
 
Grounded sample holder  SH-1.00-G, accomodates up to a Ø25 mm (1 in) sample with a  Ø25 mm (1 in)  probe area
 
Optics

     Zoom 70 monoscope      
 7:1 zoom with 9 µm resolution
     Color CCD camera  
     Swing arm  Optics can be manipulated to view any part of the sample or wafer, and can  be retracted and swung away to allow access to the top of the vacuum  chamber for sample exchange
     Video monitor  High resolution, 17-inch
     Sample illumination Coaxial via fiber optic or ring light from an adjustable light source and power supply (must specify sample illumination at time of order) NOTE: Coaxial illumination is recommended for highly reflective materials
 
High efficiency helium transfer line with foot valve for precise flow regulation
  
Basic tools, spares, and cleaning kit
 
 Home     |     Products     |     Ordering/Pricing     |     Search     |     Site Map
 