Home > Lake Shore Products > Cryogenic Probe Stations > Model CRTTP6-4K Cryogenic Probe Station Technical Specifications
 

 Model CRTTP6-4K Cryogenic Probe Station  Now available

 
Technical Specifications Model CRTTP6-4K Cryogenic Probe Station Product Overview Request A Quote for Model CRTTP6-4K Cryogenic Probe Station Downloads for Model CRTTP6-4K Cryogenic Probe Station
 
Specifications
 
Temperature*
Operating temperature range—overall     
 Base 4.5 K; control range 5 K to 475 K**
2 arms
 4 arms
 6 arms
        ZN50 DC/RF probe with low RF cryogenic coax cable
 4.5 K; 5 K to 475 K**
 5.25 K; 5.75 K to 475 K**
 6 K; 6.5 K to 475 K**
        ZN50 DC/RF probe with high RF semirigid coax cable
 4.5 K; 5 K to 400 K**
 5.25 K; 5.75 K to 400 K**
 6 K; 6.5 K to 400 K**
        GSC microwave probe with semirigid coax cable
 4.5 K; 5 K to 350 K
 5.25 K; 5.75 K to 350 K
 6 K; 6.5 K to 350 K
 
 Cool down time

        Room temperature to within 5 K of base
 1.5 h *Typical temperature performance at 60 Hz, 208 VAC, 25° C cooling water
** Selectable equipment; high temperature coaxial
sample holder limits low end temperature to 50 K
(base), 55 K (control)
        Room temperature to base  2 h
 
 Warm up time

       From base to room temperature
 1 h    
 
 Control stability

        < 10 K
 50 mK    
        11 K to 250 K  10 mK    
        251 K to 350 K  20 mK    
        351 K to 475 K**  50 mK    
 
 Temperature control (heaters)

        Sample stage
 100 W    
        Cold head  100 W    
        Radiation shield  100 W    
        CCR Riser  200 W (2 parallel heaters,  100 W each)    
        Probe arm  Measurement only    
Spacer Spacer    
CCR Vibration  
Overall (X, Y, and Z axes)
 <1 μm peak to peak at sample stage
 
Pump Down Time
Before CCR start-up (10-3 torr)
  20 min
 
Vacuum
Volume
  8.7 L (530 in3)
Room temperature
  10-4 torr
At base temperature
 10-4 torr
 
Probe Tip Temperature Drift
 
 77 K to 300 K
 Base temp to 300 K
 X axis
 TBD
 TBD
Y axis
 <5 μm
 <5 μm
 Z axis
 TBD
 TBD

 

Probe Arms  
 Travel

 X axis

 51 mm (2 in)
 Y axis  25 mm (1 in)
 Z axis  18 mm (0.7 in)
 
 Translation resolution
 X axis
 20 μm
          Y and Z axes  10 μm
          Theta planarization (included
          with microwave probes)
 ±5°
   
Sample Space  
 Sample area  Up to 51 mm (2 in)
 Sample thickness  Up to 12 mm (thicker samples may  reduce probable area)
 Working height of sample  89 cm (35 in) from floor
 
 Distance to radiation shield  4.2 cm (1.65 in) from sample holder  to bottom of viewport
 Distance from vacuum  chamber  7.2 cm (2.84 in) from sample holder
 to top of viewport
CCR Compressor Requirements
 Compressor
CSW-71C
CSW-71D
          Ambient temperature
 5 °C to 28 °C; 28 °C to 35 °C with reduced capacity
 5 °C to 28 °C; 28 °C to 35 °C with reduced capacity
          Line voltage
 208 VAC ±10%
 380/400/415 VAC
 460/480 VAC
          Frequency
 50/60 Hz
 50 Hz
 60 Hz
          Phase
 3-phase, delta
 3-phase, delta
          Power
 8.3 kW max
 8.5 kW max
 
 Compressor

          Temperature


 4 °C to 28 °C
          Flow rate 7 L/min at 28 °C to 4 L/min at 4 °C
          Inlet pressure 0.20 MPa (29 psig) to 0.69 MPa (100 psig)
          Pressure drop 0.05 MPa (7 psig) at 4 L/min to 0.32 MPa (46 psig) at 10 L/min
 
 Gas line length 6 m (19.7 ft) standard
 
 Approvals CE, UL
 
Frequency Range
ZN50 DC/RF probe frequency range
          Tungsten with cryogenic coax cable  0 to 50 MHz*
          Tungsten with semirigid coax cable  0 to 1 GHz*†
          Paliney 7 with cryogenic coax cable  0 to 50 MHz*
          Paliney 7 with semirigid coax cable  0 to 1 GHz*†
          BeCu with cryogenic coax cable  0 to 50 MHz*
          BeCu with semirigid coax cable  0 to 1 GHz*†
Spacer Spacer
GSG microwave probe frequency range
          Low frequency with K connector  0 to 40 GHz*
          Mid frequency with 2.4 mm connector  0 to 50 GHz*
          High frequency with 1.8 mm connector  0 to 67 GHz*
garrow Click here for probe & cable features
Spacer Spacer
*Selectable equipment
S21 > -10 dB up to 1 Ghz, except for a (-40 dB) spike between 400 MHz and 800 MHz depending on probe model and placement; S11 < -3 dB up to 1 GHz
 
Optical
Optical viewport—located on top lids
 Ø54 mm (2.1 in) outer window and Ø51 mm (2 in) inner window
          Outer, clear fused quartz  99% IR transmittance
          Inner
 IR absorbing with narrow band visible light transmittance
Spacer Spacer
Optical resolution—monoscope
          7:1 zoom  5 µm
          12.5:1 zoom  4 µm*
          16:1 zoom
 3 µm*
Spacer Spacer
*Selectable equipment
 
Sample Holder (Chuck)
Maximum sample size—overall  Up to Ø51 mm (2 in) *
          SH-1.25-G, Grounded chuck  Up to Ø31.8 mm (1.25 in) and 350 K
          SH-1.25-I, Isolated chuck  Up to Ø31.8 mm (1.25 in) and 350 K*
          SH-1.25-C, Coaxial chuck  Up to Ø31.8 mm (1.25 in) and 350 K*
          SH-1.25-T, Triaxial chuck  Up to Ø31.8 mm (1.25 in) and 350 K*
          SH-2.00-G, Grounded chuck  Up to Ø51 mm (2 in) and 350 K*
          SH-2.00-C, Coaxial chuck  Up to Ø51 mm (2 in) and 350 K*
          SH-2.00-C-HT, Coaxial chuck, high temperature  Up to Ø51 mm (2 in) and 475 K**
          SH-2.00-T, Triaxial chuck  Up to Ø51 mm (2 in) and 350 K*
garrow Click here to select a sample chuck (holder)  
*Selectable equipment
**Selectable equipment; high temperature coaxial sample holder limits low end temperature to 50 K (base); 55 K (control); requires option PS-HTA-CRTTP6 (purchased separately)
 
Standard Equipment
Closed cycle refrigerator — SHI RDK-408D2
 4.5 K (base); 5 K to 475 K (control)
     Sample stage temperature sensor
 Lake Shore Model DT-670-SD-1.4H calibrated silicon diode
     Cold head stage temperature sensor
 Lake Shore Model DT-670C-CU silicon diode
     Riser temperature sensor
 Lake Shore Model DT-670C-CU silicon diode
     Sample stage heater
 100 W
     Cold head stage heater
 100 W
     Riser heater
 200 W (2 parallel heaters, 100 W each)
 
Cooled radiation shield and cooled IR-absorbing window above the sample
     Diameter
 13 cm (5 in)
     Radiation shield temperature sensor
 Lake Shore Model DT-670C-CU silicon diode
     Material
 Aluminum
     Radiation shield heater
 100 W
     Removable top lid with viewpoint
 Ø51 mm (2 in) window; Ø50 mm (1.97 in) viewing area
     Sample access
 13 cm (5 in)
     Typical temperature
 30 K to 35 K throughout the full-scale temperature range
 
Temprature controllers Two Lake Shore Model 340s, one 3462 2-channel input card, and one Model 142 power supply (independent regulation of sample stage, cold head stage, riser, radiation shield, and probe arm temperature monitoring)
  
Electroless nickel plated aluminum vacuum chamber
     Diameter
 20.3 cm (8 in)
     Seals
 O-ring (370 Buna-N)
     Removable top lid with clear fused quartz viewport
 Ø63.5 mm (2.5 in) window
     Viewable area
 Ø54 mm (2.1 in)
     Probe ports
 Up to 6 surround the sample thermal radiation shield
     Pump port
 KF-40 with valve (pump sold separately)
     Gauge port
 KF-25 (gauge not included)
     Gas port
 0.25 in NPT with valve (inert gas only)
     Over pressure safety
 2 kPa (0.3 psi) pressure relief valve
 
Machined aluminum base plate (table)  63.0 cm × 50.8 cm (25 in × 20 in)
 
Support stand Low vibration, welded steel stand with integrated passive vibration isolation and dampening—minimizes system vibration displacement at all frequencies
 
Up to six XYZ precision micro-manipulated probing stages
  
Probe arms, thermal radiation shields, stainless steel welded bellows, and feedthrough ports
  
Thermally linked probe mounts  Probe mounts and radiation shield are thermally anchored to the cold head;  one probe mount includes a DT-670 series silicon diode temperature sensor  installed and wired to a 6-pin feedthrough
  
Grounded sample holder  SH-1.25-G, accommodates up to a Ø31.8 mm (1.25 in) sample with a Ø25.4  mm (1 in) probe area to land all probes
  
Optics

     Zoom 70 monoscope      
 7:1 zoom with 5 µm resolution
     Color CCD camera  S-video or composite output format
     Swing arm  Optics can be manipulated to view any part of the sample or wafer, and can  be retracted and swung away to allow access to the top of the vacuum  chamber for sample exchange
     Video monitor  High resolution, 17-inch
     Sample illumination Coaxial via fiber optic or ring light from an adjustable light source and power supply (must specify sample illumination at time of order) NOTE: Coaxial illumination is recommended for highly reflective materials
  
Instrument console
  
Basic tools, spares, and cleaning kit