Sign In
Typical Magnetic Field-Dependent Temperature Errors ΔT/T (%) at B (magnetic induction)
Range of Use
Type E thermocouple2
T (K) 2.5 T 8 T 14 T
10 1 3 7
20 <1 2 4
45 <1 <1 2
2 Useful when T ≥ 10 K. Refer to comments for Chromel-AuFe (0.07%)
Minimum
limit
Maximum limit3
Type E 3.15 K 953 K
Type K 3.15 K 1543 K
3 Upper limit dependent on wire size; to achieve higher than 473 K, insulation must be removed