Navigate Up
Sign In
Specifications
 
 
Standard curve Not applicable
 
Nominal capacitance 6.1 nF
 
Nominal sensitivity 26 pF/K
 
Accuracy (interchangeability)
Not applicable
 
Accuracy (calibrated) Calibration should be performed in situ
 
Recommended excitation 1 to 5 kHz, 0 to 7 V peak to peak or any other acceptable capacitance measuring method
 
Dissipation at recommended excitation
Not applicable
 
Expected long-term stability ±1.0 K/yr
 
Thermal response time Minutes, dominated by electronic setting time
 
Radiation effects Not available
 
Magnetic fields See table on right
 
Reproducibility Click here for detailed discussion
Range of Use
Minimum
limit
Maximum limit
CS-501GR 1.4 K 290 K

 

Click to Enlarge

Typical Magnetic Field-Dependent Temperature Errors1 ΔT/T (%) at B (magnetic induction)
Package parallel to field B
T (K) 18.7 T
4.2 -0.15
77 <0.05
1 Recommended for control purposes;
monotonic in C vs T to nearly room temperature frequency dependent
Physical Specifications
Dimensions Mass Lead type Internal
atmosphere
CS-501GR
3.0 mm ×
8.5 mm long
260 mg 2 phosphor bronze with heavy build polyimide attached with epoxy strain relief at sensor Air