| General Measurement |
Number of inputs: 3
Update rate: Up to 4 rdg/s on display; up to 18 rdg/s with IEEE-488 interface
Measurement modes: DC, RMS, peak
Probe compatibility: Standard, multi-axis, and custom probes
Probe features: Linearity Correction, Temperature Correction, Auto Probe Zero
Measurement features: Autorange, Max Hold, Relative Mode, Filter, Vector Magnitude, Differential Reading
Probe connector: 15-pin D style |
| |
| DC Measurement |
| Probe Type Ranges |
5¾-digits
with filter |
4¾-digits
without filter |
| HST Probe |
| 300 kG |
0.001 kG |
0.01 kG |
| 30 kG |
0.0001 kG |
0.001 kG |
| 3 kG |
0.00001 kG |
0.0001 kG |
| 300 G |
0.001 G |
0.01 G |
| HSE Probe |
| 30 kG |
0.0001 kG |
0.001 kG |
| 3 kG |
0.00001 kG |
0.0001 kG |
| 300 G |
0.001 G |
0.01 G |
| 30 G |
0.0001 G |
0.001 G |
| UHS Probe |
| 30 G |
0.0001 G |
0.001 G |
| 3 G |
0.00001 G |
0.0001 G |
| 300 mG |
0.001 mG |
0.01 mG |
 |
 |
 |
|
DC accuracy: ±0.10% of reading ±0.005% of range
DC temperature coefficient: ±0.05% of reading ±0.003% of range per ºC |
| |
| AC RMS & Peak Measurement |
| AC display resolution: 4¾ digits |
| Probe Type Ranges |
RMS Resolution |
Peak Resolution |
| HST Probe |
| 300 kG |
0.01 kG |
0.01 kG |
| 30 kG |
0.001 kG |
0.001 kG |
| 3 kG |
0.0001 kG |
0.0001 kG |
| 300 G |
0.01 G |
× |
| HSE Probe |
| 30 kG |
0.001 kG |
0.001 kG |
| 3 kG |
0.0001 kG |
0.0001 kG |
| 300 G |
0.01 G |
0.01 G |
| 30 G |
0.001 G |
× |
| UHS Probe |
| 30 G |
0.001 G |
0.001 G |
| 3 G |
0.0001 G |
0.0001 G |
| 300 mG |
0.01 mG |
× |
 |
 |
 |
|
AC frequency range: 10 Hz to 400 Hz
AC RMS accuracy: ±2% of reading (50 Hz to 60 Hz)
AC RMS frequency response: 0 to -3.5% of reading (10 Hz to 400 Hz)
(All AC RMS specifications for sinusoidal input >1% of range)
AC peak accuracy: ±5% typical
AC peak speed: 5 ms for single peak |
|
| Front Panel |
Display type: 4-line × 20-character, vacuum fluorescent
Display resolution: Up to ±5¾ digits
Display update rate: 4 rdg/s with Vector off, 3 rdg/s with Vector on
Display units: Gauss (G), Telsa (T)
Units multipliers: µ, m k
Annunciators: RMS: AC input signal, DC: DC input signal, MAX: Max Hold value, s: Relative reading, R: Remote operation, :Alarm on
Keypad: 25 full travel keys
Front panel features: Intutive operation, display prompts, front panel lockout, brightness control |
| |
| Interfaces |
| RS-232C capabilities |
| |
Baud: 300, 1200, 9600 |
| |
Connector: RJ-11 configuration |
| |
Update rate: Up to 14 rdg/s at 9600 baud |
 |
 |
|
| IEEE-488 capabilities |
| |
Complies with IEEE-488.2 SH1, AH1, SR1, RL1, PP0, DC1, DT0, C0, E1 |
| |
Update rate: 18 rdg/s with Vector off, 14 rdg/s with Vector on |
 |
 |
|
| Alarm |
| |
Settings: High and low setpoint, Inside/Outside, Audible |
| |
Actuators: Display annuciator, beeper |
 |
 |
|
| Monitor analog output (3) |
| |
Configuration: Real time analog voltage output |
| |
Scale: ±3 V = ±FS on selected range |
| |
Frequency response: DC to 400 Hz |
| |
Accuracy: Probe dependent |
| |
Minimum load resistance: 1 kW (short circuit protected) |
| |
Connector: BNC |
 |
 |
|
| Corrected analog output (1) |
| |
Configuration: Voltage output generated by DAC |
| |
Range: ±3 V; ±10 V for the Model 460-10 |
| |
Scale: User-defined |
| |
Resolution: 0.366 mV of ±3 V |
| |
Update rate: Same as field measurement |
| |
Accuracy: ±0.1% full scale in addition to measurement error |
| |
Minimum load resistance: 1 kW (short circuit protected) |
| |
Connector: BNC |
 |
 |
|
| |
| General |
Ambient temperature: 15 to 35 ºC at rated accuracy. 5 to 40 ºC with reduced accuracy
Power requirement: 100, 120, 220, 240 VAC (+5%, -10%), 50 or 60 Hz, 40 VA
Size: 432 mm W × 89 mm H × 368 mm D (17 in × 3.5 in × 14.5 in), full rack
Weight: 7.5 kg (16.5 lb)
Approval: CE mark |
| |
|