New Lake Shore Model 455 DSP Gaussmeter is replacing Model 450.
 
General Measurement
Number of inputs: 1

Update rate: Up to 5 rdg/s on display; up to 18 rdg/s with IEEE-488 interface

Measurement modes: DC, RMS, peak

Probe compatibility: Standard and custom probes

Probe features: Linearity Correction, Temperature Correction, Auto Probe Zero

Measurement features: Autorange, Max Hold, Relative Mode, Filter

Probe connector: 15-pin D style
 
DC Measurement
DC display resolution: 5¾ digits with filter, 4¾ digits without filter
Probe Type Ranges
Resolution
with filter
Resolution
without filter
HST Probe
300 kG
0.001 kG
0.01 kG
30 kG
0.0001 kG
0.001 kG
3 kG
0.00001 kG
0.0001 kG
300 G
0.001 G
0.01 G
HSE Probe
30 kG
0.0001 kG
0.001 kG
3 kG
0.00001 kG
0.0001 kG
300 G
0.001 G
0.01 G
30 G
0.0001 G
0.001 G
UHS Probe
30 G
0.0001 G
0.001 G
3 G
0.00001 G
0.0001 G
300 mG
0.001 mG
0.01 mG
DC accuracy: ±0.10% of reading ±0.005% of range

DC temperature coefficient: ±0.05% of reading ±0.003% of range per ºC
DC precision: No filter - ranges ≤30 G ±0.013%, ranges ≥300 G ±0.003%;
                       Filter1 - ranges ≤30 G ±0.0025%, ranges ≥300 G ±0.0007%;
1 Filter set averaging 8 readings over 1.6 s
 
AC RMS & Peak Measurement
AC display resolution: 4¾ digits
Probe Type Ranges
RMS Resolution
Peak Resolution
HST Probe
300 kG
0.01 kG
0.01 kG
30 kG
0.001 kG
0.001 kG
3 kG
0.0001 kG
0.0001 kG
300 G
0.01 G
×
HSE Probe
30 kG
0.001 kG
0.001 kG
3 kG
0.0001 kG
0.0001 kG
300 G
0.01 G
0.01 G
30 G
0.001 G
×
UHS Probe
30 G
0.001 G
0.001 G
3 G
0.0001 G
0.0001 G
300 mG
0.01 mG
×
AC RMS frequency range: 10 Hz to 400 Hz

AC RMS accuracy:
±2% of reading (50 Hz to 60 Hz)

AC RMS frequency response:
0 to -3.5% of reading (10 Hz to 400 Hz)
(All AC RMS specifications for sinusoidal input >1% of range)

AC peak accuracy: ±5% typical

AC peak speed:
5 ms
Front Panel
Display type: 2-line × 20-character, vacuum fluorescent

Display resolution:
To ±5¾ digits

Display update rate:
5 rdg/s

Display units:
Gauss (G), Telsa (T)

Units multipliers:
µ, m k

Annunciators:
RMS: AC input signal, DC: DC input signal, MAX: Max Hold value, s: Relative reading, R: Remote operation,:Alarm on

Keypad: 21 full travel keys

Front panel features: Intutive operation, display prompts, front panel lockout, brightness control
 
Interfaces
RS-232C capabilities
  Baud: 300, 1200, 9600
  Connector: RJ-11
  Update rate: Up to 15 rdg/s
IEEE-488 capabilities
  Complies with IEEE-488.2 SH1, AH1, SR1, RL1, PP0, DC1, DT0, C0, E1
  Software support : LabView™ driver
  Update rate: Up to 18 rdg/s
Alarm
  Settings: High and low set point, Inside/Outside, Audible
  Actuators: Display annuciator, beeper
Monitor analog output
  Configuration: Real time analog voltage output
  Scale: ±3 V = ±FS on selected range
  Frequency response: DC to 400 Hz
  Accuracy: Probe dependent
  Minimum load resistance: 1 kW (short circuit protected)
  Connector: BNC
Corrected analog output
  Configuration: Voltage output generated by DAC
  Range: ±3 V; ±10 V for the Model 450-10
  Scale: User defined
  Resolution: 0.366 mV
  Update rate: 5 rdg/s on display; up to 18 rdg/s with IEEE-488 interface
  Accuracy: ±0.1% full scale in addition to measurement error
  Minimum load resistance: 1 kW (short circuit protected)
  Connector: BNC
 
General
Ambient temperature: 15 to 35 ºC at rated accuracy; 5 to 40 ºC with reduced accuracy

Power requirement:
100, 120, 220, 240 VAC (+5%, -10%), 50 or 60 Hz, 20 VA

Size: 216 mm W × 89 mm H × 318 mm D (8.5 in × 3.5 in × 12.5 in), half rack

Weight: 3 kg (6.6 lb)

Approval: CE mark
 