Contact: info@lakeshore.com Lakeshore.com Search www.lakeshore.com Contact Us Ordering / Pricing Products Home
Add To Cart View Cart
Home > Lake Shore Products > Electromagnets
 Electromagnets & Recirculating Chillers
 
 
Typical Field Uniformity (EM4 Magnet)
Magnet Configuration
Uniformity
over 1 cm3
1% Cylindrical Volume
Pole Cap
cm (in)
Air Gap
cm (in)
Diameter
cm (in)
Length
cm (in)
10 (4)
5 (2.0)
±0.15%
1.8 (0.7)
5.0 (2.0)
10 (4)
2.5 (1.0)
±0.05%
6.5 (2.5)
2.5 (1.0)
7.6 (3)
2.5 (1.0)
±0.06%
4.6 (1.8)
2.5 (1.0)
5 (2)
2.5 (1.0)
±0.35%
1.8 (0.7)
2.5 (1.0)
5 (2)
1.3 (0.5)
±0.16%
3.6 (1.4)
1.2 (0.5)
The third column gives uniformity over one cubic centimeter volume centered in the magnet cap. The last two columns give the cylindrical volume within which the magnetic field deviates by less than 1% from the central field. The cylindrical volume is coaxial with the magnet poles and centered in the gap.
Typical Field Uniformity (EM7 Magnet)
Pole Cap
cm (in)
Air Gap
cm (in)
Uniformity
15.2 (6)
5 (2)
Better than ±0.1% over centered 2 in diameter circle
7.6 (3)
3.8 (1.5)
Approximately ±0.1% over centered 1 in diameter circle
 
EM4-HVA
EM7-HV
Air gap
Up to 11 cm (4.3 in) continuously variable
Up to 17.8 cm (7 in) continuously variable
Nominal coil resistance
0.25 W per coil (0.5 W wired in series)
1.0 W per coil (0.5 W wired in parallel)
Standard pole cap diameter
5.1 cm (2 in)
7.6 cm (3 in)
Optional pole cap diameter
2.5 cm, 7.6 cm, 10.2 cm (1 in, 3 in, 4 in)
15.2 cm (6 in)
Cooling water
Tap water or closed cooling system
Water flow rate
7.6 L/min (2 gal/min)
11.4 L/min (3 gal/min)
Pressure drop
200 kPa (30 psi)
220 kPa (32 psi)
Water chiller cooling capacity
2.5 kW (8,530 BTU/h)
5.0 kW (17,060 BTU/h)
Water inlet temperature
15 ºC to 25 ºC (59 ºF to 77 ºF)
15 ºC to 32 ºC (59 ºF to 89 ºF)
Coil over temperature limit
70 ºC (158 ºF)
Coil spacing, nominal
12 cm (4.75 in)
17.8 cm (7 in)
Coil size-width, nominal
12 cm (4.75 in)
13.2 cm (5.2 in)
Coil size-diameter, nominal
31.1 cm ( 12.25 in)
44.5 cm (17.5 in)
Current (maximum continuous operating)
±70 A per coil
±50 A per coil
Voltage
±35 V nominal
±50 V nominal
Maximum continuous input power
2.5 kVA nominal
5 kVA nominal
Lake Shore power supply (suggested)
Model 642
Model 665
 
Size
Height
38.7 cm at 40º angle (15.25 in) without stand
94 cm (37 in)
Width
84.5 cm (33.25 in)
101.6 cm (40 in)
Depth
50.8 cm (20 in)
66 cm (26 in)
Weight
201.9 kg (445 lb)
614 kg (1350 lb)
Shipping weight
215.5 kg (475 lb)
660 kg (1500 lb)
Shipping dimensions
95.5 cm × 58 cm × 56 cm
(38 in × 23 in × 22 in)
86 cm × 122 cm × 119.4 cm
(34 in × 48 in × 47 in)
 
Click here for Electromagnet Field Plots
Click here for Electromagnet Schematics
 
 Support Structures
EM4-HSTAND-3 - Aluminum 3-level support structure for EM4-HVA magnets
  • Bottom access to the center of the magnet
  • Compact 86.4 cm (34 in) high, 3-level aluminum structure supports the magnet and provides a flexible structure for mounting of addtional hardware
  • Leveling feet
  • Raises the center of the magnet 61 cm (2 ft) above the ground to provide a comfortable height for sample exchange
  • Plate below and above the magnet provided to support test structures or instrumentation
 
Weight: 54.4 kg (120 lb, not including magnet)

Dimensions: 58.4 cm W × 50.8 cm D × 86.4 cm H (23 in W × 20 in D × 34 in H)

Shipping weight: 269.9 kg (595 lb, with EM4-HVA magnet)

Shipping dimensions: 91.4 cm W × 61 cm D × 109.2 cm H (36 in W × 24 in D × 43 in H)
EM4-HSTAND-3
Middle plate has U-shaped slot available for bottom access
 
EM4-STAND-2 - 2-level support structure for all EM4 magnets
  • Compact 53.3 cm (21 in) high, 2-level aluminum support structure supports the magnet and provides a flexible mounting structure
  • Leveling feet
 

Weight: 35.4 kg (78 lb, not including magnet)

Dimensions: 58.4 cm W × 50.8 cm D × 53.3 cm H (23 in W × 20 in D × 21 in H)

Shipping weight: 252 kg (555 lb, with EM4-HVA magnet)

Shipping dimensions: 91.4 cm W × 61 cm D × 76.2 cm H (36 in W × 24 in D × 30 in H)

 
 Recirculating Chillers Technical Specifications
 
 
RC-EM4-200230-60-CE
RC-EM4-200-50-CE
RC-EM4-230-50-CE
RC-EM7-200230-60-CE
RC-EM7-200-50-CE
RC-EM7-400-50-CE
Electromagnet
EM4
EM4
EM4
EM7
EM7
EM7
NesLab model number
M-75
M-75
M-75
M-150
M-150
M-150
Capacity
2410 W
2180 W
2180 W
5045 W
4540 W
4540 W
Power requirements
200 to 230 V,
60 Hz, 9 A
200 V,
50 Hz, 9 A
220 to 230 V,
50 Hz, 9 A
200 to 230 V,
60 Hz, 15 A
200 V,
50 Hz, 15 A
400 V/3-phase,
50 Hz, 8 A
Volume
3.7 gpm at 40 psi
3 gpm at 40 psi
3 gpm at 40 psi
5 gpm at 45 psi
3.2 gpm at 30 psi
3.2 gpm at 30 psi
 
General
Pump
PD-2
PD-2
PD-2
CP-55
PD-2
PD-2
Temperature range
5 ºC to 35 ºC
5 ºC to 35 ºC
5 ºC to 35 ºC
5 ºC to 35 ºC
5 ºC to 35 ºC
5 ºC to 35 ºC
Temperature stability
±0.15 ºC
±0.15 ºC
±0.15 ºC
±0.15 ºC
±0.15 ºC
±0.15 ºC
Reservoir volume (gal/L)
0.5/1.8
0.5/1.8
0.5/1.8
0.5/1.8
0.5/1.8
0.5/1.8
Case dimensions (in) (cm)
26.2 × 16.3 × 24
66.4 × 41.3 × 61.0
26.2 × 16.3 × 24
66.4 × 41.3 × 61.0
26.2 × 16.3 × 24
66.4 × 41.3 × 61.0
30.4 × 21.3 × 29.3
77.2 × 54× 74.3
30.4 × 21.3 × 29.3
77.2 × 54× 74.3
30.4 × 21.3 × 29.3
77.2 × 54× 74.3
Shipping weight
91 kg (200 lb)
91 kg (200 lb)
91 kg (200 lb)
130 kg (287 lb)
130 kg (287 lb)
130 kg (287 lb)
Performance specifications will be affected by changes in temperature, ambient, or coolant.
* M-150 can be used with the EM7 magnet, but do not operate the magnet above the rated wattage.
 
 
 
 Home     |     Products     |     Ordering/Pricing     |     Search     |     Site Map
 
Privacy Policy | ©2007 Lake Shore Cryotronics, Inc.
 
Product Overview Ordering Information Downloads